INTRODUCTION | ||||||||||||
In recent years, the capabilities of ultrasound flow imaging have
increased enormously. Color flow imaging is now commonplace and
facilities such as ‘power’ or ‘energy’ Doppler
provide new ways of imaging flow. With such versatility, it is tempting
to employ the technique for ever more demanding applications and
to try to measure increasingly subtle changes in the maternal and
fetal circulations. To avoid misinterpretation of results, however,
it is essential for the user of Doppler ultrasound to be aware of
the factors that affect the Doppler signal, be it a color flow image
or a Doppler sonogram.
|
||||||||||||
Competent use of Doppler ultrasound techniques requires an understanding
of three key components:
(1) The capabilities and limitations of Doppler ultrasound;
(2) The different parameters which contribute to the flow display; (3) Blood flow in arteries and veins.
This chapter describes how these components contribute to the quality
of Doppler ultrasound images. Guidelines are given on how to obtain
good images in all flow imaging modes. For further reading on the
subject, there are texts available covering Doppler ultrasound and
blood flow theory in more detail 1-3 .
|
||||||||||||
BASIC PRINCIPLES | ||||||||||||
Ultrasound images of flow, whether color flow or spectral Doppler,
are essentially obtained from measurements of movement. In ultrasound
scanners, a series of pulses is transmitted to detect movement of
blood. Echoes from stationary tissue are the same from pulse to
pulse. Echoes from moving scatterers exhibit slight differences
in the time for the signal to be returned to the receiver (Figure
1 ). These differences can be measured as a direct
time difference or, more usually, in terms of a phase shift from
which the ‘Doppler frequency’ is obtained (Figure
2). They are then processed to produce either a color
flow display or a Doppler sonogram.
|
||||||||||||
As
can be seen from Figures
1 and
2, there has to be motion in the direction of the
beam; if the flow is perpendicular to the beam, there is no relative
motion from pulse to pulse. The size of the Doppler signal is
dependent on:
(1) Blood velocity: as velocity increases, so does the Doppler
frequency;
(2) Ultrasound frequency: higher ultrasound frequencies give increased Doppler frequency. As in B-mode, lower ultrasound frequencies have better penetration. (3) The choice of frequency is a compromise between better sensitivity to flow or better penetration; (4 The angle of insonation: the Doppler frequency increases as the Doppler ultrasound beam becomes more aligned to the flow direction (the angle q between the beam and the direction of flow becomes smaller). This is of the utmost importance in the use of Doppler ultrasound. The implications are illustrated schematically in Figure 3. |
||||||||||||
All
types of Doppler ultrasound equipment employ filters to cut out
the high amplitude, low-frequency Doppler signals resulting from
tissue movement, for instance due to vessel wall motion. Filter
frequency can usually be altered by the user, for example, to exclude
frequencies below 50, 100 or 200 Hz. This filter frequency limits
the minimum flow velocities that can be measured.
|
||||||||||||
CONTINUOUS
WAVE AND PULSED WAVE
|
||||||||||||
As
the name suggests, continuous wave systems use continuous transmission
and reception of ultrasound. Doppler signals are obtained from all
vessels in the path of the ultrasound beam (until the ultrasound
beam becomes sufficiently attenuated due to depth). Continuous wave
Doppler ultrasound is unable to determine the specific location
of velocities within the beam and cannot be used to produce color
flow images. Relatively inexpensive Doppler ultrasound systems are
available which employ continuous wave probes to give Doppler output
without the addition of B-mode images. Continuous wave Doppler is
also used in adult cardiac scanners to investigate the high velocities
in the aorta.
|
||||||||||||
|
||||||||||||
Doppler
ultrasound in general and obstetric ultrasound scanners uses pulsed
wave ultrasound. This allows measurement of the depth (or range)
of the flow site. Additionally, the size of the sample volume (or
range gate) can be changed. Pulsed wave ultrasound is used to provide
data for Doppler sonograms and color flow images.
|
||||||||||||
Aliasing | ||||||||||||
Pulsed
wave systems suffer from a fundamental limitation. When pulses are
transmitted at a given sampling frequency (known as the pulse repetition
frequency), the maximum Doppler frequency fd that can be measured
unambiguously is half the pulse repetition frequency. If the blood
velocity and beam/flow angle being measured combine to give a fd
value greater than half of the pulse repetition
frequency, ambiguity in the Doppler signal occurs. This ambiguity
is known as aliasing. A similar effect is seen in films where wagon
wheels can appear to be going backwards due to the low frame rate
of the film causing misinterpretation of the movement of the wheel
spokes.
|
||||||||||||
|
||||||||||||
The
pulse repetition frequency is itself constrained by the range of
the sample volume. The time interval between sampling pulses must
be sufficient for a pulse to make the return journey from the transducer
to the reflector and back. If a second pulse is sent before the
first is received, the receiver cannot discriminate between the
reflected signal from both pulses and ambiguity in the range of
the sample volume ensues. As the depth of investigation increases,
the journey time of the pulse to and from the reflector is increased,
reducing the pulse repetition frequency for unambiguous ranging.
The result is that the maximum fd
measurable decreases with depth.
Low
pulse repetition frequencies are employed to examine low velocities
(e.g. venous flow). The longer interval between pulses allows the
scanner a better chance of identifying slow flow. Aliasing will
occur if low pulse repetition frequencies or velocity scales are
used and high velocities are encountered (Figure
4,5 and 6). Conversely, if a high pulse repetition
frequency is used to examine high velocities, low velocities may
not be identified.
|
||||||||||||
|
||||||||||||
ULTRASOUND FLOW MODES | ||||||||||||
Since
color flow imaging provides a limited amount of information over
a large region, and spectral Doppler provides more detailed information
about a small region, the two modes are complementary and, in practice,
are used as such.
Color
flow imaging can be used to identify vessels requiring examination,
to identify the presence and direction of flow, to highlight gross
circulation anomalies, throughout the entire color flow image, and
to provide beam/vessel angle correction for velocity measurements.
Pulsed wave Doppler is used to provide analysis of the flow at specific
sites in the vessel under investigation. When using color flow imaging
with pulsed wave Doppler, the color flow/B-mode image is frozen
while the pulsed wave Doppler is activated. Recently, some manufacturers
have produced concurrent color flow imaging and pulsed wave Doppler,
sometimes referred to as triplex scanning.
When
these modes are used simultaneously, the performance of each is
decreased. Because transducer elements are employed in three modes
(B-mode, color flow and pulsed wave Doppler), the frame rate is
decreased, the color flow box is reduced in size and the available
pulse repetition frequency is reduced, leading to increased susceptibility
to aliasing.
Power Doppler is also referred to as energy Doppler, amplitude Doppler and Doppler angiography. The magnitude of the color flow output is displayed rather than the Doppler frequency signal. Power Doppler does not display flow direction or different velocities. It is often used in conjunction with frame averaging to increase sensitivity to low flows and velocities. It complements the other two modes (Table 01). Hybrid color flow modes incorporating power and velocity data are also available from some manufacturers. These can also have improved sensitivity to low flow. A brief summary of factors influencing the displays in each mode is given in the following sections. Most of these factors are set up approximately for a particular mode when the application (e.g. fetal scan) is chosen, although the operator will usually alter many of the controls during the scan to optimize the image. |
||||||||||||
|
||||||||||||
|
||||||||||||
Color flow imaging | ||||||||||||
Color
flow Doppler ultrasound produces a color-coded map of Doppler shifts
superimposed onto a B-mode ultrasound image (Color
Flow Maps). Although color flow imaging uses pulsed
wave ultrasound, its processing differs from that used to provide
the Doppler sonogram. Color flow imaging may have to produce several
thousand color points of flow information for each frame superimposed
on the B-mode image. Color flow imaging uses fewer, shorter pulses
along each color scan line of the image to give a mean frequency
shift and a variance at each small area of measurement. This frequency
shift is displayed as a color pixel. The scanner then repeats this
for several lines to build up the color image, which is superimposed
onto the B-mode image. The transducer elements are switched rapidly
between B-mode and color flow imaging to give an impression of a
combined simultaneous image. The pulses used for color flow imaging
are typically three to four times longer than those for the B-mode
image, with a corresponding loss of axial resolution.
Assignment
of color to frequency shifts is usually based on direction (for
example, red for Doppler shifts towards the ultrasound beam and
blue for shifts away from it) and magnitude (different color hues
or lighter saturation for higher frequency shifts). The color Doppler
image is dependent on general Doppler factors, particularly the
need for a good beam/flow angle. Curvilinear and phased array transducers
have a radiating pattern of ultrasound beams that can produce complex
color flow images, depending on the orientation of the arteries
and veins. In practice, the experienced operator alters the scanning
approach to obtain good insonation angles so as to achieve unambiguous
flow images.
|
||||||||||||
FACTORS AFFECTING THE COLOR FLOW IMAGE | ||||||||||||
The
controls that affect the appearance of the color flow image are
summarized in Table
2. The main factors include:
(1) Power and gain:Color flow uses higher-intensity
power than B-mode. Attention should be paid to safety indices. Power
and gain should be set to obtain good signal for flow and to minimize
the signals from surrounding tissue.
(2)
Frequency selection: Many scanner/transducer
combinations permit changes of frequency. High frequencies give
better sensitivity to low flow and have better spatial resolution.
Low frequencies have better penetration (Figure 5) and are less
susceptible to aliasing at high velocities.
(3)
Velocity scale/pulse repetition frequency: Low
pulse repetition frequencies should be used to examine low velocities
but aliasing may occur if high velocities are encountered (Figura
7a,b).
(4)
Region of interest: Because more pulses are needed
to look at flow than for the B-mode image, reducing the width and
maximum depth of the color flow area under investigation will usually
improve frame rate and may allow a higher color scan line density
with improved spatial resolution (Figure
9).
(5) Focus: The focus should be at the level
of the area of interest. This can make a significant difference
to the appearance and accuracy of the image (Figure 7).
In
practice, the operator will make many changes to the controls and
will try different probe positions to optimize the image. Practical
guidelines are given in Table
3.
|
||||||||||||
SPECTRAL OR PULSED WAVE DOPPLER | ||||||||||||
Pulsed
wave Doppler ultrasound is used to provide a sonogram of the artery
or vein under investigation (Figure
12). The sonogram provides a measure of the changing
velocity throughout the cardiac cycle and the distribution of velocities
in the sample volume (or gate) (Figure
11). If an accurate angle correction is made, then
absolute velocities can be measured. The best resolution of the
sonogram occurs when the B-mode image and color image are frozen,
allowing all the time to be employed for spectral Doppler. If concurrent
imaging is used (real-time duplex or triplex imaging), the temporal
resolution of the sonogram is compromised.
|
||||||||||||
FACTORS AFFECTING THE SPECTRAL IMAGE | ||||||||||||
The
controls that affect the appearance of the sonogram are summarized
in Table 4.
The main factors include:
(1) Power and gain: Pulsed wave Doppler
uses higher intensity power than B-mode. Attention should be paid
to safety indices. Power and gain should be set so that clear signals
are obtained.
(2) Velocity scale/pulse repetition frequency:
Low pulse repetition frequencies should be used to look at low
velocities but aliasing may occur if high velocities are encountered.
(3) Gate size: If flow measurements are
being attempted, the whole vessel should be insonated. A large gate
may include signals from adjacent vessels (Figure
13).
|
||||||||||||
|
||||||||||||
Guidelines for a practical approach to obtain good-quality spectral images are given in Table 5 . | ||||||||||||
|
||||||||||||
BLOOD FLOW MEASUREMENTS | ||||||||||||
Velocity measurement | ||||||||||||
Theoretically,
once the beam/flow angle is known, velocities can be calculated
from the Doppler spectrum as shown in the Doppler equation. However,
errors in the measured velocity may still occur 1,4.
Sources of error can be broadly divided into three categories.
(1) Errors can arise in the formation of the Doppler
spectrum due to:
(a) Use of multiple elements in array transducers;
(b) Non-uniform insonation of the vessel lumen;
(c) Insonation of more than one vessel;
(d) Use of filters removing low-velocity components.
(2) Errors can arise in the measurement of the
ultrasound beam/flow velocity angle.
(a)
Use of high angles (q > 60o) may give rise to error
because of the comparatively large changes in the cosine of the
angle which occur with small changes of angle (Figure
14).
(b)
The velocity vector may not be in the direction of the vessel axis.
(3)
Errors can arise in the calculation packages provided by the manufacturers
for analysis of the Doppler spectrum (for instance, of intensity
weighted mean velocity).
(a)
While efforts can be made to minimize errors, the operator should
be aware of their likely range. It is good practice to try to repeat
velocity measurements, if possible using a different beam approach,
to gain a feel for the variability of measurements in a particular
application. However, even repeated measurements may not reveal
systematic errors occurring in a particular machine.
(b)
The effort applied to produce accurate velocity measurements should
be balanced against the importance of absolute velocity measurements
for an investigation.
(c)
Changes in velocity and velocity waveform shape are often of more
clinical relevance when making a diagnosis. In this and other cases,
absolute values of velocity measurement may not be required.
|
||||||||||||
|
||||||||||||
Calculation of absolute flow | ||||||||||||
Total
flow measurement using color or duplex Doppler ultrasound is fraught
with difficulties, even under ideal conditions 5. Errors
that may arise include:
(1) Those due to inaccurate measurement of vessel cross-sectional
area, for example the cross-sectional area of arteries which pulsate
during the cardiac cycle;
(2) Those originating in the derivation of velocity (see above).
These errors become particularly large when flow calculations are
made in small vessels; errors in measurement of diameter are magnified
when the diameter is used to derive cross-sectional area. As with
velocity measurements, it is prudent to be aware of possible errors
and to conduct repeatability tests.
|
||||||||||||
Flow waveform analysis | ||||||||||||
Non-dimensional
analysis of the flow waveform shape and spectrum has proved to be
a useful technique in the investigation of many vascular beds. It
has the advantage that derived indices are independent of the beam/flow
angle.
Changes in flow waveform shape have been used to investigate both
proximal disease (e.g. in the adult peripheral arterial circulation)
and distal changes (in the fetal circulation and uterine arteries).
While the breadth of possible uses shows the technique to be versatile,
it also serves as a reminder of the range of factors which cause
changes to the local Doppler spectrum. If waveform analysis is to
be used to observe changes in one component of the proximal or distal
vasculature, consideration must be given to what effects other components
may have on the waveform.
|
||||||||||||
Flow waveform shape: indices of measurement | ||||||||||||
Many
different indices have been used to describe the shape of flow waveforms
1 . Techniques range from simple indices of systolic
to diastolic flow to feature extraction methods such as principal
component analysis. All are designed to describe the waveform in
a quantitative way, usually as a guide to some kind of classification.
In general, they are a compromise between simplicity and the amount
of information obtained.
|
||||||||||||
|
||||||||||||
The
relative merits of indices used in uterine arteries have been discussed
elsewhere. 6,7 .Commonly used indices available on most
commercial scanners are:
(1) Resistance index (RI) (also called resistive index or Pourcelot’s
index);
(2) Systolic/diastolic (S/D) ratio, sometimes called the A/B ratio;
(3) Pulsatility index (PI) 8.
These indices are all based on the maximum Doppler shift waveform
and their calculation is described in Figure 12. The PI takes slightly
longer to calculate than the RI or S/D ratio because of the need
to measure the mean height of the waveform. It does, however, give
a broader range of values, for instance in describing a range of
waveform shapes when there is no end-diastolic flow.
|
||||||||||||
|
||||||||||||
In
addition to these indices, the flow waveform may be described or
categorized by the presence or absence of a particular feature,
for example the absence of end-diastolic flow and the presence of
a post-systolic notch.
Generally, a low pulsatility waveform is indicative of low distal
resistance and high pulsatility waveforms occur in high-resistance
vascular beds (Figure 8), although the presence of proximal stenosis,
vascular steal or arteriovenous fistulas can modify waveform shape.
Care should be taken when trying to interpret indices as absolute
measurements of either upstream or downstream factors. For example,
alterations in heart rate can alter the flow waveform shape and
cause significant changes in the value of indices.
|
Thursday, 10 May 2012
DOPPLER ULTRASOUND
Subscribe to:
Post Comments (Atom)
No comments:
Post a Comment